Potentiometry passively measures the potential of a solution between two electrodes, affecting the solution very little in the process. One electrode is called the reference electrode and has a constant potential, while the other one is an indicator electrode whose potential changes with the sample's composition. Therefore, the difference in ...
Reference electrodes generally used are hydrogen electrodes, calomel electrodes, and silver chloride electrodes. The indicator electrode forms an electrochemical half-cell with the interested ions in the test solution. The reference electrode forms the other half-cell. The overall electric potential is calculated as
An ion-selective electrode (ISE), also known as a specific ion electrode (SIE), is a simple membrane-based potentiometric device which measures the activity of ions in solution. [1] It is a transducer (or sensor ) that converts the change in the concentration of a specific ion dissolved in a solution into an electrical potential .
Whether the analyte is reduced or oxidized depends on the analyte and the potential applied, but its reaction always occurs at the working/indicator electrode. Therefore, the working electrode potential varies as a function of the analyte concentration. A second auxiliary electrode completes the electric circuit, called the counter electrode.
The connection can be direct, through a narrow tube to reduce mixing, or through a salt bridge, depending on the other electrode and solution. This creates an ionically conductive path to the working electrode of interest. A reference electrode is an electrode that has a stable and well-known electrode potential.
Commercial reference electrodes consist of a glass or plastic tube electrode body. The electrode consists of a metallic silver wire (Ag (s)) coated with a thin layer of silver chloride (AgCl), either physically by dipping the wire in molten silver chloride, chemically by electroplating the wire in concentrated hydrochloric acid (HCl) [3] or electrochemically by oxidising the silver at an anode ...
The auxiliary electrode may be isolated from the working electrode using a glass frit. Such isolation prevents any byproducts generated at the auxiliary electrode from contaminating the main test solution: for example, if a reduction is being performed at the working electrode in aqueous solution, oxygen may be evolved from the auxiliary electrode.
The signal is measured as the potential difference (voltage) between the working electrode and the reference electrode. The working electrode's potential must depend on the concentration of the analyte in the gas or solution phase. The reference electrode is needed to provide a defined reference potential.